ID102840 Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.
Symptom
Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76
Error message
3602019 Bad kV Detected DO XRTechMgr acquisition Component(s) _Tube 1 0 [XRTechMgr::notifyGeneratorFinalExposureParameters] numberOfTubeSpits=X above the maxNumberOfTubeSpits=X after X frames XRTechMgr.cpp 1964 3602019 Bad kV Detected DO XRTechMgr acquisition Component(s) _Tube 1 0 [XRTechMgr::notifyGeneratorFinalExposureParameters] numberOfTubeSpits=X above the maxNumberOfTubeSpits=X after X frames XRTechMgr.cpp 1964 3602085 Jedi: too many tube spits Data Related to Error: X (Anode + Cathode Spits), AR 3602085 Jedi: too many tube spits AR XRTechMgr acquisition Jedi too many tube spits 10 JEDI Error Code: 0316H Spit Max error Data Related to Error: X (Anode + Cathode Spits) Data related to Error codes: 4,5,6,10. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service". 3602085 Jedi: too many tube spits AR XRTechMgr acquisition Jedi too many tube spits 10 JEDI Error Code: 0316H Spit Max error Data Related to Error: X (Anode + Cathode Spits). Data related to Error codes: 4,5,6,10 Review the Data related to the error. If Anode + Cathode Spits Data related to error = 4 select this concept. Data Related to error indicates the spit type: Error:1 = Anode - Gro
Applies to
- Models
- Innova
- Product
- Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- Ulysses MX160 X-Ray Tube Ulysses MX160 X-Ray Tube Part number 2216500 (new) or D2802A (reload) Ulysses MX160 X-Ray Tube Part number 2216500 (new) or D2802A (reload)
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.