GE Healthcare · MR

OrthoMRI DPA Hand/Wrist coil (coil 6) is below SNR specifications.OrthoMRI Poor image quality on wrist scans.OrthoMRI Re…

OrthoMRI DPA Hand/Wrist coil (coil 6) is below SNR specifications.OrthoMRI Poor image quality on wrist scans.OrthoMRI Replaced the DPA hand/wrist coil (coil 6).OrthoMRI Cscan

Symptom

OrthoMRI Poor image quality on wrist scans.

Applies to

Product
OrthoMRI Cscan
Software
OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
Component
OrthoMRI LNR9101643000 DPA Hand/Wrist coil See long description. When ordering the coil, the system center frequency must be added to the comments of the order for these coils are pre-tuned by the manufacturer. Once received, RF180 and varicap calibrations must be completed prior to customer use.
Area
MR

Full cause & step-by-step resolution