Pular para o conteúdo principal
GE Healthcare · MRExcite224601411134978042247814222385010303200002352

System crashes when attempting to scan due to corrupt software.

Symptom

Can Not Scan

Error message

2246014 Operating system software error Error: 2246014 Host: ghmr001 Process: diagCoord File: dcCoordWrite.c Line: 179 Task ID: 0x0 Time: 1113497804 Object: Diag Coordinator Operating system software error.

2247814 Diagnostic Communications (DC) error. Error when reading socket connection 2247814 Host: GEMS Process: diagCoord File: procDcMsg.c Line: 521 Diagnostic Communications (DC) error. Error when reading socket connection. An automatic attempt will be made to recover. File descriptor: 12 Object: Host Diag Executive Error code: -1-

2223850 Content Of Packet Cant Be Routed To A Host Application Thu Feb 2 11:01:58 2006 Error: 2223850 Host: RIMIMR Process: NSP File: PMH.cc Line: 456 Contents of packet can't be routed to a host application: opcode:10303 seq:20 rev:2 type:1 length:8 stat:-1 resp:WavRec req:HC

200002352 A HOST_SHUTDOWN was initiated by the death of scn

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required