Cardiac MR-Echo Scans Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" Due To SRF/TRF…
Cardiac MR-Echo Scans Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" Due To SRF/TRF Board
Symptom
Cardiac MR-Echo Scans Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" See Long Description System "locks up" or "hangs" intermittently during Cardiac MR-Echo scans, and logs the error messages shown below. CSO 1-1156077914, iTrak 13169945. Mon Feb 11 15:08:33 2008 2248008 4 WCB_MR1_OC NSP APS:DRF 250 Exception Class: Abort Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: MR-Echo Coil: 8CARDIAC Scan: 8598/14 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: 1 FE ID: 0 segCount: 0 frameCount: 4653070 Mon Feb 11 15:08:34 2008 2254907 1 WCB_MR1_OC LxMxConsumer gemrimxframework.appcontroller.scanner.ge 0 Scan Failed.
Error message
2254907 Scan Failed. 2254907 WCB_MR1_OC LxMxConsumer gemrimxframework.appcontroller.scanner.ge 0 Scan Failed. 2248008 RFM EOS semaphore timeout error. IRF did not send enough data to DRF.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 2280951 SRF/TRF board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.