Cardiac MR-Echo Scans Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" Due To SRF/TRF…

Cardiac MR-Echo Scans Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" Due To SRF/TRF Board

Symptom

Cardiac MR-Echo Scans Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" See Long Description System "locks up" or "hangs" intermittently during Cardiac MR-Echo scans, and logs the error messages shown below. CSO 1-1156077914, iTrak 13169945. Mon Feb 11 15:08:33 2008 2248008 4 WCB_MR1_OC NSP APS:DRF 250 Exception Class: Abort Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: MR-Echo Coil: 8CARDIAC Scan: 8598/14 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: 1 FE ID: 0 segCount: 0 frameCount: 4653070 Mon Feb 11 15:08:34 2008 2254907 1 WCB_MR1_OC LxMxConsumer gemrimxframework.appcontroller.scanner.ge 0 Scan Failed.

Error message

2254907 Scan Failed. 2254907 WCB_MR1_OC LxMxConsumer gemrimxframework.appcontroller.scanner.ge 0 Scan Failed.

2248008 RFM EOS semaphore timeout error. IRF did not send enough data to DRF.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
2280951 SRF/TRF board
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.