Scanning has stopped Master Gradient is not ready due to burned HFA capacitor.
Symptom
Scans Abort with Gradient Amplifier % Axis Undervoltage and Gradient Fidelity Faults See Long Description for error string. Mon Jan 14 14:14:23 2008 2244570 2 MRI_15T_PBOI NSP SCP:Grad 1099 The Z Axis gradient amplifier failed to go to Ready when commanded.-> Mon Jan 14 14:14:23 2008 2244577 2 MRI_15T_PBOI NSP SCP:Grad 2796 The GP is reporting a Gradient Amplifier Z Axis UnderVoltage Fault.-> Mon Jan 14 14:14:25 2008 2218527 1 MRI_15T_PBOI scn MsaScanOpsScrn.m 916 Scanning has stopped. Master Gradient is not ready. Mon Jan 14 14:14:25 2008 2218624 1 MRI_15T_PBOI scn StateMachine.c 200 Failed to execute the next event.
Error message
2244850 The GP is reporting a Gradient Amplifier X Axis UnderVoltageFault See long description 2244850 the GP is reporting a Gradient Amplifier X Axis UnderVoltageFault 2244570 The gradient amplifier failed to go to Ready when commanded. 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.-> 2244570 TOiMR2 NSP SCP:Grad 1106 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 The X Axis gradient amplifier failed to go to Ready when commanded.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- SecondOpinion: HFA with FRU Package and Lift Brackets 2393906-4
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.