Pular para o conteúdo principal
GE Healthcare · MRExcite

Can the M3078JJ Medrad HD NV array coil be installed on a 11x, Excite2 scanner.No, the M3078JJ Medrad HD NV array coil i…

Can the M3078JJ Medrad HD NV array coil be installed on a 11x, Excite2 scanner.No, the M3078JJ Medrad HD NV array coil is for Excite HD 12.0 and Higher Platforms.

Symptom

Can the M3078JJ Medrad HD NV array coil be installed on a 11x, Excite2 scanner? Site wishes to use the HD NV array coil from the sister MR site which is an upgraded 14.0 system.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
M3087JJ Medrad 1.5T HD Neurovascular Array
Area
MR

Full cause & step-by-step resolution