GE Healthcare · MR23001
OrthoMRI Incorrect magnet temperature.OrthoMRI Cannot run a scout.OrthoMRI Replaced the SCTERM unit.OrthoMRI Escan (Unix…
OrthoMRI Incorrect magnet temperature.OrthoMRI Cannot run a scout.OrthoMRI Replaced the SCTERM unit.OrthoMRI Escan (Unix Digital).OrthoMRI 23001 N.
Symptom
OrthoMRI Cannot run a scout OrthoMRI Cannot run a scout.
Error message
OrthoMRI 23001 N: Incorrect temperature, one or more temperatures aren't in the required range.
Applies to
- Product
- OrthoMRI Escan (Unix Digital) Two easy ways to tell if the Escan Unix system is analog or digital. First, the analog service password is coil and the digital password is probe. Second, the analog daily QA phantom is a bottle where as the digital daily QA phantom is a large spherical ball.
- Software
- OrthoMRI Unix 2.x
- Component
- OrthoMRI LNR9101148101 SCTERM Unit See long description. You must insert the correct SCTERM offset and the room temperature offset (read them from the label on the front side of the new SCTERM). Remember that the offsets are important to correctly read the magnet temperature errors but they don’t have any influence on the magnet warm up. The room temperature offset can be set only using step equal to ten units (e.g.: 10, 20, 30,……., 60, 70, etc)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- OrthoMRI Network cable is unplugged
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.