OrthoMRI Incorrect magnet temperature.OrthoMRI Cannot run a scout.OrthoMRI Replaced the SCTERM unit.OrthoMRI Escan (Unix…

OrthoMRI Incorrect magnet temperature.OrthoMRI Cannot run a scout.OrthoMRI Replaced the SCTERM unit.OrthoMRI Escan (Unix Digital).OrthoMRI 23001 N.

Symptom

OrthoMRI Cannot run a scout OrthoMRI Cannot run a scout.

Error message

OrthoMRI 23001 N: Incorrect temperature, one or more temperatures aren't in the required range.

Applies to

Product
OrthoMRI Escan (Unix Digital) Two easy ways to tell if the Escan Unix system is analog or digital. First, the analog service password is coil and the digital password is probe. Second, the analog daily QA phantom is a bottle where as the digital daily QA phantom is a large spherical ball.
Software
OrthoMRI Unix 2.x
Component
OrthoMRI LNR9101148101 SCTERM Unit See long description. You must insert the correct SCTERM offset and the room temperature offset (read them from the label on the front side of the new SCTERM). Remember that the offsets are important to correctly read the magnet temperature errors but they don’t have any influence on the magnet warm up. The room temperature offset can be set only using step equal to ten units (e.g.: 10, 20, 30,……., 60, 70, etc)
Area
MR

Full cause & step-by-step resolution