8 Channel HR Brain Coil validation error Due to CoilConfig.cfg corrupted.

Symptom

1.5T HD 8 Channel HR Brain Coil Validation Error 1.5T HD 8 Channel HR Brain Coil Validation Error.

Error message

2258113 Host coil validation error: Default Config Bias(defaultBiasCfg) field not found in CoilConfig.cfg. Connector: Multicoil Lower Port Coil: 8NVHEAD_A. Pls call your service representative 2258113 TRMR01 coilCK coilCK_scp.c 910 Host coil validation error: Default Config Bias(defaultBiasCfg) field not found in CoilConfig.cfg. Connector: Multicoil Lower Port Coil: 8NVHEAD_A. Pls call your service representative

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 M5 Excite HDx HDi Platforms only.
Area
MR