GE Healthcare · X-ray VascularInnova501201050088

Innova: Positioner error.

Positioner error.

Symptom

Innova: Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited, Lost All Gantry axis's drive or NO movements. Omega table elevation will not drive down. Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited Positioner error. Reset system. If problem persists, call Service. X-Ray is inhibited, Lost All Gantry axis's drive or NO movements. Omega table elevation will not drive down. Lift failure, dose is minimized, call service

Error message

50120 PIB: ENABLE, HARDSTOP, BUMPER OR PRD 50120 "POS : PIB: ENABLE, HARDSTOP, BUMPER OR PRD ERROR " See for HW code 50120 "POS : PIB: ENABLE, HARDSTOP, BUMPER OR PRD ERROR " DO POS 0 0 HWStatus=0017 User=0000 0002 4AC5 XRTitan.cpp 50120: POS:PIB: ENABLE, HARDSTOP, BUMPER OR PRD ERROR: DO: HWStatus=0019 User=0000 0002 4AC5

INX100: 1050088 PIB INTERNAL WR GantryCPU No Boot PIB_INTERNAL 0 0 - RELAY_K12, open,Stuck 1050088 PIB INTERNAL WR GantryCPU No Boot PIB_INTERNAL 0 0 - RELAY_K12, Open or Stuck. The Pib has detected relay K12 is open or Stuck or monitoring error. 1050088 PIB INTERNAL WR GantryCPU No Boot PIB_INTERNAL 0 0 - RELAY_K12, open,Stuck, or monitoring error. 1050088 PIB INTERNAL WR GantryCPU No Boot PIB_INTERNAL 0 0 - RELAY_K12, Open or Stuck. The Pib has detected relay K12 is open or Stuck or monitoring error. Relay K12 is used for checking the emergency backout + Hardstop circuit of the lift. The test of this relay is Hardstop + emergency backout drive.

INX100: 1050088 WR GantryCPU PIB INTERNAL Debug: - RELAY_K1, open

INX100: 1050088 WR GantryCPU PIB INTERNAL Debug: - RELAY_K14, open 1050088 WR GantryCPU PIB INTERNAL Debug: - RELAY_K14, open. This error can be a perm

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems
Component
Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.