Loose cable between WIM and SCIM.

Symptom

The PDU is in Stop Mode and the EMO Reset Will Not Work See Long Description Sometimes when you push the green EMO reset button on the PDU nothing happens. You cannot hear the 24VDC relay click on nor hear any fans start up. In addition, you keep getting an error message that the PDU is in stop mode or the 24VDC supply has failed after you reset the TPS and attempt to scan.

Error message

2248579 A timeout occurred when trying to receive a SCIM packet 2248579 Host: PMR5_OC0 Process: NSP File: SCP:KeyboardLinkResources Line: 471 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Gradient Echo Coil: HEAD Scan: 28727/1 SCIM Communication Error. A timeout occurred when trying to receive a SCIM packet. Opcode = 0x60 Check cable connection between the SCIM and WIM and cable connection between the SCIM and System Cabinet.

2248592 SCIM Configuration Failed SCIM link errors caused the configuration to fail 2248592 Host: PMR5_OC0 Process: NSP File: SCP:OpWorkspaceAutoConfig Line: 364 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Gradient Echo Coil: HEAD Scan: 28727/1 SCIM Configuration Failed. SCIM link errors caused the configuration to fail. Please fix and perform a TPS Reset

2248602 PDU in Stop Mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2248602 PDU is in Stop mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed.

Applies to

Models
Excite, Signa
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.