GE Healthcare · MRExcite2242664
ECG and Cardiac Gating have no Signal Due to Defective ECG extention cable.
Symptom
ECG and Cardiac Gating have no Signal Cannot change from peripheral gating to ECG gating (any lead), no signal. ECG and Cardiac Gating have no Signal
Error message
2242664 Unable to Find a Valid ECG Trigger for 10 Secs. Error: 2242664 Host: lx-mr Process: NSP File: AGP:wait for valid gate(): No ECG Line: 0 Timed out. Unable to find a valid ECG trigger for 10 secs.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 2219096 ECG EXTENSION CABLE MG2A11-ECG
- Area
- MR