GE Healthcare · MRExcite225425667103
Troubleshooting 3T MKS RF Amplifier PA Air Flow fault 41.
Symptom
Troubleshooting 3T MKS RF Amplifier PA Air Flow fault 41 Troubleshooting 3T RF Amplifier PA Air Flow fault 41
Error message
2254256 3T RF Amplifier PA Air Flow (fault 41) see long description Sun Mar 18 17:42:13 2007 Error: 2254256 Host: MR0C1 Process: NSP File: SCP:../rf_nb_contro Line: 1491 3T RF Amplifier PA Air Flow (fault 41). Possible Causes: dirty air filters, bad fuses, defective sensor board. Check that air filters are clean. Test fuses on the Power Supply deck. If the fault persists, replace the RF amplifier.-> NOTE: This error can be cause by one of the 3 "hidden fuses" opening. See SN 67103 for details.
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 2336064-8 3T CORONA MKS RF AMPLIFIER FUSE KIT 3T CORONA RF AMPLIFIER FUSE KIT 2336064-8 3T CORONA RF AMPLIFIER FUSE KIT 2336064-8
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.