Alarm — stator control board error conditionGantry Rotation Control

ER0DTorque voltage measurement/output tolerance exceeded

Philips CT error code, reported by the Gantry Rotation Control (Stator Control Board).

What ER0D means

On Access Series CT (6/16/Dual) CT systems, status code ER0D indicates: Torque voltage measurement/output tolerance exceeded. The code originates in the Gantry Rotation Control section of the Gantry Rotation Control (Stator Control Board). Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5506. Per the guide, the tolerance between the measured and output torque voltage exceeded the set value; follow the Gantry Rotation Prepare Error flow chart (Section 2.4.5) and check the servo torque feedback wiring before board replacement..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.

Full troubleshooting for ER0D

The complete diagnosis, cause tree, and step-by-step remedy for the Gantry Rotation Control subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Gantry Rotation Control codes

Discuss ER0D with other field engineers in the FSELIB forum

Source: Philips Access CT Service Trouble Shooting Guide — 459800863401 Rev. C. Reference information re-authored by FSELIB — not a copy of the OEM manual.