Philips CT error code, reported by the Gantry Rotation Control (Stator Control Board).

What ER09 means

On Access Series CT (6/16/Dual) CT systems, status code ER09 indicates: Gantry rotor motor servo warning. The code originates in the Gantry Rotation Control section of the Gantry Rotation Control (Stator Control Board). Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5502 in the Error Log Viewer/system log (Section 2.9), then follow the Gantry Rotation Prepare Error flow chart (Section 2.4.5). Check the rotor motor servo drive and its wiring before replacing the Stator Control Board..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.