Alarm — stator control board error conditionGantry Rotation Control

ER0ARotation pulse sampling abnormal

Philips CT error code, reported by the Gantry Rotation Control (Stator Control Board).

What ER0A means

On Access Series CT (6/16/Dual) CT systems, status code ER0A indicates: Rotation pulse sampling abnormal. The code originates in the Gantry Rotation Control section of the Gantry Rotation Control (Stator Control Board). Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5503 in the system log. Follow the Gantry Rotation Prepare Error flow chart (Section 2.4.5); check the rotation pulse sensor/cable and Rotor Control Board (see V26 'Rotating pulse counting' and V32 'Rotating pulse cable' LEDs in Table 1-1)..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.