Alarm — stator control board error conditionGantry Rotation Control

ER0CGantry stop-rotation time-out

Philips CT error code, reported by the Gantry Rotation Control (Stator Control Board).

What ER0C means

On Access Series CT (6/16/Dual) CT systems, status code ER0C indicates: Gantry stop-rotation time-out. The code originates in the Gantry Rotation Control section of the Gantry Rotation Control (Stator Control Board). Severity level alarm (Alarm — stator control board error condition).

Reported action: Cross-reference System Error Code 0x5505 in the system log. Follow the Gantry Rotation Prepare Error flow chart (Section 2.4.5) and check the rotor motor servo braking/stop response before replacing the Stator/Rotor Control Board..

Listed in Table 1-3 'Stator Control Board Error Code vs System Error Code' of the Philips Access CT Service Trouble Shooting Guide (459800863401 Rev. C), for the Access Series CT (6/16/Dual) family. Not confirmed for other Philips CT systems.