245180 — Gantry display blanks after an OBC LAN watchdog reset during AutoMA
GE Healthcare CT error code, reported by the GE LightSpeed CT platform.
What 245180 means
On LightSpeed QX/i and LightSpeed Plus and LightSpeed Ultra and LightSpeed 16 and HiSpeed QXi CT systems, status code 245180 indicates: Gantry display blanks after an OBC LAN watchdog reset during AutoMA. The code originates in the Table/Gantry — OBC Firmware section of the GE LightSpeed CT platform. Severity level error (Error condition).
Reported action: Read this together with error 245043: it is a known LS2002 M4 firmware defect, not a hardware fault, so do not start swapping Gantry/OBC parts. A hardware reset from the Utilities screen clears it, and dropping AutoMA from the protocol is the documented workaround pending the firmware fix. Details are in the service manual..
Raised on GE LightSpeed QX/i, Plus, Ultra, LightSpeed 16 and HiSpeed QXi systems running LS2002 M4 Table/Gantry firmware — the same AutoMA/OBC LAN watchdog fault documented for error 245043 from this KO: on the last image of a series with AutoMA on, the OBC resets on a bad AutoMA table-size command and logs a LAN watchdog error that blanks the Gantry display.
Full troubleshooting for 245180
The complete diagnosis, cause tree, and step-by-step remedy for the Table/Gantry — OBC Firmware subsystem are documented in the FSELIB GE Healthcare service manual — re-authored, web-first, with real figures.
7-day free trial · no credit card required
Related Table/Gantry — OBC Firmware codes
Source: GE Offline Knowledge (kOffline). Reference information re-authored by FSELIB — not a copy of the OEM manual.