200302183System fails to restore calibration data (cals) from stateMOD

GE Healthcare CT error code, reported by the GE LightSpeed CT platform.

What 200302183 means

On LightSpeed QX/i and LightSpeed Plus and LightSpeed Ultra and LightSpeed 16 and CT/i CT systems, status code 200302183 indicates: System fails to restore calibration data (cals) from stateMOD. The code originates in the OCTANE Computer section of the GE LightSpeed CT platform. Severity level error (Error condition).

Reported action: This ties back to the Octane console's IP30 board being heat-sensitive: the related MOD/archive faults on the same console were traced to internal temperature climbing above roughly 82F. Check the console's cooling and airflow before assuming a bad disk or a corrupted state file — the full IP30 diagnostic and board-swap procedure is in the service manual..

Reported by the sysStateABB process (ss_osdep.c) on the Octane console of LightSpeed QX/i, LightSpeed Plus, LightSpeed Ultra, LightSpeed 16 and CT/i systems when the console cannot restore calibration data (cals) from the archived stateMOD. Logged in the same GE field-service solution that also documents Octane MOD-archive attach failures (errors 10102 and 10180) on this console, both traced to the Octane-I IP30 board's sensitivity to internal temperature.

Full troubleshooting for 200302183

The complete diagnosis, cause tree, and step-by-step remedy for the OCTANE Computer subsystem are documented in the FSELIB GE Healthcare service manual — re-authored, web-first, with real figures.