Error conditionTable/Gantry — OBC Firmware

245043OBC LAN watchdog error blanks the Gantry display during an AutoMA scan

GE Healthcare CT error code, reported by the GE LightSpeed CT platform.

What 245043 means

On LightSpeed QX/i and LightSpeed Plus and LightSpeed Ultra and LightSpeed 16 and HiSpeed QXi CT systems, status code 245043 indicates: OBC LAN watchdog error blanks the Gantry display during an AutoMA scan. The code originates in the Table/Gantry — OBC Firmware section of the GE LightSpeed CT platform. Severity level error (Error condition).

Reported action: This is a known LS2002 M4 firmware defect, not a hardware failure — do not start swapping Gantry or OBC parts. A simple hardware reset from the Utilities screen clears it, and avoiding AutoMA in the protocol is the documented workaround until the M5 (or 4.x Linux/GRE M3) firmware fix. Details are in the service manual..

Raised on GE LightSpeed QX/i, Plus, Ultra, LightSpeed 16 and HiSpeed QXi systems running LS2002 M4 Table/Gantry firmware: on the last image of a series with AutoMA turned on, the OBC gets an AutoMA table-size command it can't handle correctly, resets, and logs a LAN watchdog error (STC Master Exc Handler, exm_lan_wdog.c) that blanks the Gantry display and stops scanning.

Full troubleshooting for 245043