Error conditionScan Data Acquisition (SDAS) — Data Disk

245047Scan stops mid-acquisition due to a scan data disk chain fault

GE Healthcare CT error code, reported by the GE LightSpeed CT platform.

What 245047 means

On LightSpeed QX/i and LightSpeed Ultra and LightSpeed Plus CT systems, status code 245047 indicates: Scan stops mid-acquisition due to a scan data disk chain fault. The code originates in the Scan Data Acquisition (SDAS) — Data Disk section of the GE LightSpeed CT platform. Severity level error (Error condition).

Reported action: Read this together with error 230013015 — pull both the gesys_ct01.log and the iceConsole.log and work the scan-data-disk chain (cable, controller, RIP board, power, fans) rather than guessing at a single part. The full check sequence and a real-case resolution are in the service manual..

Raised on GE LightSpeed QX/i, Ultra and Plus CT systems when a scan stops mid-acquisition (the operator can resume it with the resume key) and both the gesys_ct01.log and the iceConsole.log point to the scan data disk chain — the same fault chain documented for error 230013015 from this KO: raw-data SCSI disk, cabling, SCSI controller board, RIP Motorola CPU board, power supplies, or cooling fans.

Full troubleshooting for 245047

The complete diagnosis, cause tree, and step-by-step remedy for the Scan Data Acquisition (SDAS) — Data Disk subsystem are documented in the FSELIB GE Healthcare service manual — re-authored, web-first, with real figures.