Pular para o conteúdo principal
Error conditionScan Data Acquisition (SDAS) — Data Disk

230013015Scan halts mid-acquisition with a SCSI scan data disk (DAS) fault

GE Healthcare CT error code, reported by the GE LightSpeed CT platform.

What 230013015 means

On LightSpeed QX/i and LightSpeed Ultra and LightSpeed Plus CT systems, status code 230013015 indicates: Scan halts mid-acquisition with a SCSI scan data disk (DAS) fault. The code originates in the Scan Data Acquisition (SDAS) — Data Disk section of the GE LightSpeed CT platform. Severity level error (Error condition).

Reported action: Pull both the gesys_ct01.log and the iceConsole.log together — neither log alone is conclusive — and work the scan-data-disk chain (cable, controller, RIP board, power, fans) before condemning a single part. This KO traced a real case to a bad raw-data disk plus a corrupted recon queue; the full check sequence is in the service manual..

Raised on GE LightSpeed QX/i, Ultra and Plus CT systems when a scan stops mid-acquisition (the operator can resume it with the resume key) and both the gesys_ct01.log and the iceConsole.log point to the scan data disk chain: the raw-data SCSI disk, its cabling, the SCSI controller board, the RIP Motorola CPU board, power supplies, or cooling fans.

Full troubleshooting for 230013015

The complete diagnosis, cause tree, and step-by-step remedy for the Scan Data Acquisition (SDAS) — Data Disk subsystem are documented in the FSELIB GE Healthcare service manual — re-authored, web-first, with real figures.

7-day free trial · no credit card required

Related Scan Data Acquisition (SDAS) — Data Disk codes

Discuss 230013015 with other field engineers in the FSELIB forum

Source: GE Offline Knowledge (kOffline). Reference information re-authored by FSELIB — not a copy of the OEM manual.