244585 — Scanning hardware reset completes after an RCIB fault-line mismatch
GE Healthcare CT error code, reported by the GE LightSpeed CT platform.
What 244585 means
On LightSpeed QX/i and LightSpeed Plus CT systems, status code 244585 indicates: Scanning hardware reset completes after an RCIB fault-line mismatch. The code originates in the SDAS (Scan Data Acquisition) section of the GE LightSpeed CT platform. Severity level error (Error condition).
Reported action: If it shows up once at power-up and the system resets cleanly on its own, it can usually be left alone; if it repeats, check that the DCB J4 cable sits on the J4 connector and the terminator sits on J5 — a swapped J4/J5 connection on the DCB is a documented cause. The full RCIB/CAN-bus wiring reference is in the service manual..
Raised on the Scan Data Acquisition (SDAS) subsystem of GE LightSpeed QX/i and Plus CT systems, typically on power-up: the DCB fails to sense the correct state of one or more RCIB fault lines (a serial/parallel fault-line mismatch), and this code — logged by fwmgr in HwManager.c — reports that the resulting scanning-hardware reset completed successfully.
Full troubleshooting for 244585
The complete diagnosis, cause tree, and step-by-step remedy for the SDAS (Scan Data Acquisition) subsystem are documented in the FSELIB GE Healthcare service manual — re-authored, web-first, with real figures.