Error conditionICE Chassis

230006014ICE chassis fails to boot after the OC loses the cu ice serial link

GE Healthcare CT error code, reported by the GE LightSpeed CT platform.

What 230006014 means

On LightSpeed QX/i and LightSpeed Plus and LightSpeed Ultra and HiSpeed QXi CT systems, status code 230006014 indicates: ICE chassis fails to boot after the OC loses the cu ice serial link. The code originates in the ICE Chassis section of the GE LightSpeed CT platform. Severity level error (Error condition).

Reported action: Treat this together with error 230006001 — check the console-to-ICE cu ice serial link and the UUCP Systems file rather than swapping ICE hardware first. Full recovery steps and the correct Systems file contents are in the service manual..

Raised on GE LightSpeed CT systems (QX/i, Plus, Ultra) and HiSpeed QXi as part of the same ICE-chassis boot failure documented for error 230006001: the console cannot cu to the ICE chassis, so the ICE fails to boot and a HOST_SHUTDOWN follows the death of the procICE process. Logged in iceCpu.cxx.

Full troubleshooting for 230006014

The complete diagnosis, cause tree, and step-by-step remedy for the ICE Chassis subsystem are documented in the FSELIB GE Healthcare service manual — re-authored, web-first, with real figures.