Pular para o conteúdo principal
AlarmFilm Tray Module — Sensitometry Calibration

HC11314Film calibration required for the film in the upper tray

Agfa CT error code, reported by the Film Tray Module.

What HC11314 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11314 indicates: Film calibration required for the film in the upper tray. The code originates in the Film Tray Module — Sensitometry Calibration section of the Film Tray Module. Severity level alarm (Alarm).

Reported action: Run a film (sensitometry) calibration for the film loaded in the upper tray..

Raised on the Drystar 5500/5503 when the printhead has been removed and replaced and the sensitometry calibration on file for the film loaded in the upper tray is no longer valid for the newly installed printhead. The printer shows "film calib. upper tray needed".

Full troubleshooting for HC11314

The complete diagnosis, cause tree, and step-by-step remedy for the Film Tray Module — Sensitometry Calibration subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Film Tray Module — Sensitometry Calibration codes

Discuss HC11314 with other field engineers in the FSELIB forum

Source: AGFA DryStar service manual — HC error catalogue (FSELIB authored corpus). Reference information re-authored by FSELIB — not a copy of the OEM manual.