Pular para o conteúdo principal
AlarmAlignment Module — Entry Sensor (S22)

HC11213Sheet did not clear alignment entry sensor S22 in time (input jam)

Agfa CT error code, reported by the Alignment Module.

What HC11213 means

On DryStar 5500 (Type 5364/100) and DryStar 5503 (Type 5364/300) CT systems, status code HC11213 indicates: Sheet did not clear alignment entry sensor S22 in time (input jam). The code originates in the Alignment Module — Entry Sensor (S22) section of the Alignment Module. Severity level alarm (Alarm).

Reported action: The source does not list a reason or cure specific to this code. Treat it as a film-path jam at sensor S22: check the print section for a jammed sheet and clear it, then verify S22 is correctly positioned and not mechanically broken; if the fault keeps recurring, inspect the alignment module for mechanical blockage of sheet transport..

Raised on the Drystar 5500/5503 when the trailing end of a sheet has not passed the vertical alignment entry-detection sensor S22 within the time-out limit, after the sheet was detected passing S22. The printer shows "Remove input jam". Related to the other S22-based codes covering the entry-detection sensor (HC10826, HC11211, HC11233, HC11234).

Full troubleshooting for HC11213

The complete diagnosis, cause tree, and step-by-step remedy for the Alignment Module — Entry Sensor (S22) subsystem are documented in the FSELIB Activion16 (TSX-031A) service manual — re-authored, web-first, with real figures.

Open the Activion16 service manual

Related Alignment Module — Entry Sensor (S22) codes

Discuss HC11213 with other field engineers in the FSELIB forum