GE Healthcare · CTLightSpeed260003164502671579950646260003149
Scanning aborted with collimator errors
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 260003164 A Z-Axis Update Contained a Move That Is Invalid For CAM A. Illegal distance for CAM A encoder count conversion. Illegal distance for CAM B encoder count conversion. See LD for complete error Fri Oct 12 12:43:36 2001 Error: 260003164 Host: HSC Process: CCB Firmware File: aperture_z_axis.c Line: 230 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50267/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:43:36:745 Log Series: 57 A Z-Axis Update Contained a Move That Is Invalid For CAM A. Position : 15799 Sequence Number: 0 Mon Oct 2 10:06:25 2006 Error: 260003164 Host: HSC Process: CCB Firmware File: aperture_z_axis.c Line: 230 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50646/4/0 Type: None/Unknown Exception Level: Pri/Soft Time: 10:06:25:445 Log Series: 898 A Z-Axis Update Contained a Move That Is Invalid For CAM B. Position : -259 Sequence Number: 38 Error: 260003149 Illegal distance for CAM A encoder count conversion. Could be Cam B also See LD for complete error codes Fri Oct 12 12:43:36 2001 Error: 260003149 Host: HSC Process: CCB Firmware File: cfg_cam.c Line: 191 Function: XRAY GENERATION : Xray Collimation and F
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required