IBXACQ system PMTUNE 4th or 5th iteration lasted over 2 minutes while it attempted to adjust the LED Rec.

Symptom

IBXACQ system PMTUNE 4th or 5th iteration lasted over 2 minutes while it attempted to adjust the LED Rec. PMTUNE program failed with message saying "Setup can not be completed for this scan" When LED Rec. adjustment was done manually from the Service menu it would display the same message (Curve is too low) but the adjustment would actually complete correctly and the E signal in the IB Box would look normal. However, it would still fail during Pmtune Update. Contributor 212002357 Nielsen March 12 2008

Error message

LED recovery failed - 6 retries. Curve is too low. Error message in the error log (listerr): "LED recovery failed - 6 retries. Curve is too low." Followed by some messages indicating the scan failed.