MRSPT BODY FSE Stability fails, only on the last 2 echos of the echo train.

Symptom

MRSPT BODY FSE Stability fails SPT, only on the last 2 echoes of the echo train. FGRE Stability is clean. see long description The system is a 1.5T LX HiSpeed with an Indigo computer, running 8.3 M5 software. SPT fails on the BODY FSE stability. Strange thing about this failure is that it only fails on last 2 echos of the echo train. I think with this kind of a failure, something is happening in the data acquisition portion in the TPS. See the attached for examples of the SPT plots. lx-mr 4# more system_health.log System Health script run on: 03/04/04 09:02:54 hostname = lx-mr Hospital Name = -------------------------------------------------- Config File Checks -------------------------------------------------- Platform type = LX System software revision = Operating System Version = 6.5 Field Strength = 15000 SR Mode = 77 Configured for broadband spectroscopy = NO Governing Body = fda2 absoluteAvgSar = 2.0 discreteAvgSar = 2.0 absolutePeakSar = 8.0 discretePeakSar = 8.0 dbdtMHW = 20 absoluteDbdtIEC = 66 discreteDbdtIEC = 20 dbdtFDA1 = 45 dbdtFDA2 = 66 Units for Dbdt research is % absoluteDbdtResearch = 66 discreteDbdtResearch = 66 Magnet type = 1.5T LCC Multi Coil = YES RF Amp type = 1.5T 1.0T Erbtec tube-type Grad Coil Type = 60cm BRM GradWarp coeff file correct...........................PASS Grad Amp Type = Techron 8645 Gradients Fast TG is enabled in /w/config/prescan_config.cfg No functions are emulated out in the ipg_stage file. Head Amp Cal Factor = 0...............................PASS Body Amp Cal Factor = 0...............................PASS psdRfWait = 156 Expected = 156........................PASS psdGradWait = 171 Expected = 172......................FAIL Grad cal values of 31689 31039 30206 all pos..........PASS Gradient rise time = 284 for this SR mode.............PASS RF amp safety limits in MR config file are correct -------------------------------------------------------

Error message

2222617 TPS IPG BOARD - Fast Update Mechanism failure. Instruction memory bank 1 has failed the address test. This bank is used for Rho, Theta, and Omega generation. see long description Error seen during TPS diags: 1094647477 2 1 Wed Sep 8 12:44:37 2004 2222617 2 lx-mr diags diagsMain.cc 2263 TPS IPG BOARD - Fast Update Mechanism failure. Instruction memory bank 1 has failed the address test. This bank is used for Rho, Theta, and Omega generation. Address: 00d80002H Data Exp: 0001H Data Rd: 0081H EN 310 SR 117

2222250 The Sequencer RAM address line test from the 68020 failed. Check the address bus or the Sequencer RAM on the IPG board. see long description Error seen with TPS diags: 1094647477 0 1 Wed Sep 8 12:44:37 2004 2222250 2 lx-mr diags diagsMain.cc 2263 IPG BOARD LEVEL DIAGS FAILURE The Sequencer RAM address line test from the 68020 failed. Check the address bus or the Sequencer RAM on the IPG board. address : 00000000H written: 00d80002H expected: 00000002H actual: 00000002H EN 330 SR 291

2222267 IPG Sequencer RAM walking 1s and 0s test failed. see long description Error seen with TPS diags: 1094647477 1 1 Wed Sep 8 12:44:37 2004 2222267 2 lx-mr diags diagsMain.cc 2263 I

Applies to

Models
Signa, HiSpeed
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
8.3 8.3
Area
MR