Pular para o conteúdo principal
GE Healthcare · MRSigna2248108

CWA ID98607. Unable to scan with DRF Over Range Errors.

Symptom

Unable to scan with DRF Over Range Errors APS received data over range errors from DRF2 on all channels (0-7), error number 2248108. Shortly after, there may or may not be an error indicating the master gradient failed to go to ready when commanded.

Error message

2248108 APS received data over range error from DRF, on chan= n. 2248108 TOiMR2 NSP APS:dataMoverTask 604 Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 APS received data over range error from DRF, on chan = 2 Need to reduce receiver gain.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR