GE Healthcare · MRSigna2225612
Will Not Scan With RF Errors due to the Processor Board.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2225612 ERBTEC RF Amplifier Fault 21. 24V power supply unacceptable Error: 2225612 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3060 ERBTEC RF Amplifier fault 21. 24V power supply unacceptable. Less than 22.3V or greater than 25V. Possible defective Power Control Board.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-305000P4 Erbtec Amplifier PROCESSOR BOARD For 1.0/1.5T Amplifier NOTE 1. CAREFULLY REMOVE U16 CHIP (68HC11) AND U17 CHIP (EPROM) FROM ITS SOCKETS BY GENTLY PRYING UP BOTH ENDS SIMULTANEOUSLY (USE AN IC REMOVING TOOL IF POSSIBLE). PROTECT AND SET ASIDE FOR REPLACEMENT IN NEW PROCESSOR BOARD. CAREFULLY REMOVE U16 CHIP (68HC11) AND U17 CHIP (EPROM) FROM ITS SOCKETS BY GENTLY PRYING UP BOTH ENDS SIMULTANEOUSLY (USE AN IC REMOVING TOOL IF POSSIBLE). PROTECT AND SET ASIDE FOR REPLACEMENT IN NEW PROCESSOR BOARD.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.