GE Healthcare · MRSigna22257372225738
Unable to scan with Power Monitor Faults due to HN Bulkhead Connector (Body Heliax Feed Through Connector).
Symptom
Unable to scan with Power Monitor Faults
Error message
2225737 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Thu Mar 7 16:22:41 2002 2225737 2 lx-mr NSP HPC_PORT_ID:spi_pm_log_error 100 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 40 pulse width: 33 duty cycle: 03 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan. This could also occur when power cycling SSM.> Thu Mar 7 16:22:41 2002 2225737 2 lx-mr NSP HPC_PORT_ID:spi_pm_log_error 100 Power Monitor 1 Fault.RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 40 pulse width: 33 duty cycle: 03 Fault in head amplitude.If you need to confirm the source of the fault, reset the power monitor and repeat the scan. This could also occur when power cycling SSM.> 2225738 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Thu Mar 7 16:22:41 2002 2225738 2 lx-mr NSP HPC_PORT_ID:spi_pm_log_error 100 Power Monitor 2 Fa
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-221942P1 UG-1019 HN BULKHEAD 50-OHM JACK-JACK This is the RF body transmit heliax HN jack-jack straight feed through connector. See Long Description for a photograph UG-1019 HN BULKHEAD 50-OHM JACK-JACK 46-221942P1
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.