Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Mouse changes state from arrow icon to hour glass icon in viewport. See Long Description for details Mouse changes state from arrow icon to hour glass icon in viewport. Intermittently the mouse changed states while in the viewports, causing havoc in trying to execute mouse clicks.

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-28 is a 2 button PS2 mouse used on GOC1 and earlier consoles (those without HP computer). Part # 2142759 is obsolete and this 3 button mouse is a replacement. It is probable that any 3 button mouse will work in place of a 2 button mouse. 2169940-28 is a 2 button PS2 mouse used on GOC1 and earlier consoles (those without HP computer). Part # 2142759 is obsolete and this 3 button mouse is a replacement. It is probable that any 3 button mouse will work in place of a 2 button mouse. Should NOT be used on GOC2 and above or GRE consoles which contain the HP computer. The HP computer uses a 3 button mouse which is part # 2368697-2 or part # 2348405-2. 2169940-28 is a 2 button PS2 mouse used on GOC1 and earlier consoles (those without HP computer). Part # 2142759 is obsolete and this 3 button mouse is a replacement. It is probable that any 3 button mouse will work in place of a 2 button mouse. Should NOT be used on GOC2 and above or GRE consoles which contain the HP computer. The HP computer uses a 3 button mouse which is part # 2368697-2 or part # 2348405-2. Part #: 2169940-28 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCTANE MOUSE FRU: Yes Status: Orderable Comments: FLS - OCTANE MOUSE & INDIGO COMPUTERU/O NEXGEN 12/97 GWENN MS08992REPLACES 2114561-12 FOR (LX) LIGHTNING. PER PCN, 197770 AND FMI#60476 MUST ORDER 2187507 FOR CONTOUR SAME AS 2142759 (ALL GONE!), WHICH WAS PHASED OUT WHEN STK.WAS USED UP. MS7093 1-7-00 MOUSE AND PAD PER CHIP NLA AS HOTLINE/OBTAINFROM V/C 627584 5/1/00 MOUSE PAD IS 2252872. FE/CUSTOMER NOTE: CT ENGINEERING HAS AUTHORIZED S.G.I. TO DELIVER EITHER LIGHTOR DARK GRAY MOUSE TO FILL ORDERS FOR THIS ITEM. MS7093 HQT: M3 NEXGEN FW 10 '98 * NO LONTER STK'D @ MR MFG'RG 4-00 FOR EXCITE II SYSTEM, USE P/N 2368697-2, PER RSPL OF SAME NAME. MS20544 8/25/03. GELS - SAME AS 2142759,OCTANE MOUSE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues