GE Healthcare · CTLightSpeedVCT2601181775005450148
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
CTx00875 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service. CTx00875 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service.
Error message
Error: 260118177 Jedi (40-0114)Rotor not at speed. See Long Description for complete error Thu Feb 15 14:07:54 2007 260118177 4 orp orp_subsys gen_exec.c 2376 Thu Feb 15 14:07:54 2007 Host : Jedi Ermes # : 260118177 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Static Scan: 50054/1/1 (40-0114) Rotor not at speed . Z6,Z7= 2h,0h Mon Apr 17 14:01:27 2006 Error: 260118177 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Mon Apr 17 14:01:26 2006 Host : Jedi Ermes # : 260118177 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Static Scan: 50148/1/1 (40-0114)Rotor not at speed. ^C Z6,Z7= 2h,0h
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Area
- CT
Full cause & step-by-step resolution