GE Healthcare · CTLightSpeed245190230000069291443700050003244000
LightSpeed 5.X (80/100kW)
Symptom
Can't perform Scout Scan. Can't perform Scout Scan Please go to the error concept and check all that apply The customer may see Scanner Hardware Stopped Scan message in the operator console area 1 Use the following script and search the message log at time of Scout failure to get the correct errors. [insite@erct ~]$ /usr/g/insite/bin/viewlog 2 Type those errors in the psdb search box separate by spaces 3 Go to the error concept and check the appropriate errors
Error message
Error: 245190 System aborted Axial scanning. No views, either offset or data, were collected/received. Data views expected: 992 See the Long Description for the full error code. Thu Jan 13 06:58:23 2005 Error: 245190 Host: ct02 Process: scanmgr File: scan_complete.cxx Line: 303 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1000 System aborted Axial scanning. No views, either offset or data, were collected/received. Data views expected: 992 Do Not Use - use one of the other Error: 230000069 Two, or more, unexpected events were detected during scan data transfer. The 230000069 Error has several varation see below. Sun Sep 5 11:10:31 2004 Error: 230000069 Host: darc Process: dataacq File: dip_thread.cxx Line: 396 Function: Data Acquisition : Scan Data Acquisition Scan Type: Scout Scan: 4167/1/1 Scan Seq Id: 1000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 3 1 TAXI modem violation 5 1 View length error Mon Dec 15 10:40:11 2008 span style="COLOR: #ff0000"> Error: 230000069 Host: darc Process: dataacq File: hddip.cxx Line: 720 Function: Data Acquisition : Scan Data Acquisition Scan Type: Helical
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P31 FUSE 1/2 AMP SB 250V See LD for more info 46-170021P31 1/2A, 250V SLO-BLO FUSE. TYPE 3AG, 1.25~ X 0.25~ GLASS BODY Part #: 46-170021P31 Part Type: GP CONSUMABLES Part Weight: 0.02 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE 1/2 AMP SB 250V FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.