Detector may see panel temperature rises that could potentially affect Image Quality Apply SN 30022
Symptom
Detector may see panel temperature rises that could potentially affect Image Quality The Detector may see panel temperature rises that could potentially affect Image Quality (if too high) or shut down operation due to a variety of fault possibilities including: 1) Conditioner Airflow: a) Inadequate airflow inside DSA to remove chiller exhaust and power supply heat b) Clogged chiller air filter c) Blockage of chiller or DSA exhaust 2) Coolant Flow: a) Kinked hoses b) Clog in coolant lines, pumps, etc 3) Detector Cold Plate: a) Inefficiency of heat transfer from cold plat
Applies to
- Models
- Definium
- Product
- Definium 8000
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required