GE Healthcare · X-ray R&FDefinium
Def8000 OTS Collimator has no power.
Definium 8000 OTS Collimator has no power due to a tripped breaker (CB8) ont he Tube Suspension Logic (TSL) Board.
Symptom
Def8000 OTS Collimator has no power. OTS Power Supply is good, fuses are good. +30 Vac to collimator on pins 7/9 and 11/15 is missing.
Applies to
- Models
- Definium
- Product
- Definium 8000
- Component
- OTS
- Area
- X-ray R&F