Pular para o conteúdo principal
GE Healthcare · CTHiSpeedLightSpeed183345

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

Scanner would not scan on large or small filament

Error message

Error: 183345 The mA board version is incorrect for tube type. See Long Description for details Mon Dec 17 20:05:35 2007 Error: 183345 Suite: RP26 Host: OBC Process: Generator File: hemrc manager.c 1.20 Method: No Method Line: 169 Function: XRAY GENERATION : MA Loop Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 20:05:30:505 Log Series: 23 The mA board version is incorrect for tube type. MA version: 46-288886G1 Tube ID: 8 (Zeus = 7, Gemini = 8) Zeus = 7 is the Glass Tube D3102T and Gemini = 8 is the Performix Tube D3132T The error above was caused by installing the older 46-288886G1 MA bd. into a system that had a Performix Gemini Tube D3132T

Applies to

Models
HiSpeed, LightSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2154834 MA Board used on (CT/i,1.X-4.X) scanners See Long Description for more Part #: 2154834 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.81 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: MA CONTROL BRD./CANCEL FRU: Yes Status: Orderable Comments: FLS - MA BOARD USED ON FAST RHAPSODE FMI#25238 UPGRADES THIS BD. TO A REV.1 ONE USE LIGHTSPEED ULTRA HQT WARNING!! MATERIAL MUST BE PURCHASED FROM MFGING DO NOT GO TO VENDOR AS IT MUST BE TESTED IN MFGING BEFORE SHIPPED TO FIELD VERY VERY HIGH DOA RATE. MS08992 GVR 7/31/00GELS
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.