LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Images report zero mA or 1999mA with noise in image or starved images See long description for picture of image and also IMPOSSIBLE ma readout Images report 0ma or 1999ma with noise in image or starved images Look at both pictures below!

Error message

Error: 260132451 The Beam Tracking Algorihtm Encountered a Division by Zero Sat Sep 17 03:13:33 2005 Error: 260132451 Host: orp Process: orp_subsys File: Tracking.cxx Line: 542 Sat Sep 17 03:13:32 2005 Host : Das Ermes # : 260132451 Exception Class : Abort Severity : Secondary File : Tracking.cxx Line# : 542 Function : No System Function Reported Scan Type : Helical Scan: 542/2/1 The Beam Tracking Algorihtm Encountered a Division by Zero.

Error: 290012000 View corrections were applied in view # The errror 290012000 "View corrections were applied in view number" message indicate the Image Generator or recon software detected a corrupt view and has applied view corrections to image. This Error alone does not indicate a problem. See service note SN05102815a. Wed Feb 13 13:46:25 2008 Error: 290012000 Host: ig3 Process: image_generation File: MissedView.cpp Line: 129 View corrections were applied in view 5400. exam/series/scan = 345/2/1 From SN05102815a Error code 290012000 is a status notification that Recon has made a correction to maintain image quality. This error is transparent to the Customer and has no effect on system performance. This condition is generated when the IG reads i

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution