Pular para o conteúdo principal
GE Healthcare · CTLightSpeed185109

94831 LightSpeed 4.X

Symptom

Cradle will not move in or out of the Gantry. No c-pulse is being detected from longitudal encoder Cradle will not move in or out of the Gantr . No c-pulse is being detected from longitudal encoder y

Error message

Error: 185109 Elevation Interference/Limit switches are INVALID. See Long Description for full error Mon Dec 10 16:56:08 2007 Error: 185109 Host: ETC Process: Control File: etcValidIntfMatrix.c 1.8 7/9/ Line: 96 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 16:56:08:425 Log Series: 45 HW error: Elevation Interference/Limit switches are INVALID. Interference/Limit switches = 0xc1d0. UP LIMIT: 1 DOWN LIMIT: 0 INTF1: 1 INTF2: 0 (0 = NOT active 1 = active)

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269173 ELEVATION ENCODER ASM FOR H2 LCC -Zeus Table see Long Description for details and picture Part #: 2269173 Part Type: GP CONSUMABLES Part Weight: 3.99 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEVATION ENCODER ASM FOR H2 LCC FRU: Yes Status: Orderable Comments: FLS - ELEVATION ENCODER ADDED. USED ON H2 TABLE LCC. ASIA HAS IDENTIFIED THIS AS A LINKED PART (SEE 46-296633G3).AWAITING CONFIRMATION ON GL18.MSA8429 11/8/01 WILL START DEPLOYMENT FROM ASIA ON FW12, 02. MS30544 1/31/02
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.