LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183126 FIL/MA fault latch 2 would not reset. See Long Description for sample error. Tue Nov 13 08:44:03 2007 Error: 183126 Host: OBC Process: Generator File: gen abort.c 1.31 Line: 901 Function: XRAY GENERATION : Filament Selection Scan: 15769/3/1 Type: Helical Exception Level: Secondary Time: 08:44:04:090 Log Series: 297 FIL/MA fault latch 2 would not reset. Hardware address = 0xffdf4d, Status = 0xc0

Error: 183123 FIL/MA fault detected: Filament short circuit. Desired focal spot = 0. (0 = large, 1 = small) See Long Description for sample error. Tue Nov 13 08:39:25 2007 Error: 183123 Host: OBC Process: Generator File: gen abort.c 1.31 Line: 376 Function: XRAY GENERATION : Filament Selection Scan: 15769/2/1 Type: Helical Exception Level: Secondary Time: 08:39:24:655 Log Series: 129 FIL/MA fault detected: Filament short circuit. Desired focal spot = 0. (0 = large, 1 = small)

Error: 183124 FIL/MA fault detected: Inoperative filament inverter. See LD for more details Mon Aug 7 08:55:55 2006 Error: 183124 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 385 Function: XRAY GENERATION : Filament Selection Scan: 849/2/6 Type: Helical Exception Level: Pri/Most Severe Time:

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2154834 MA Board used on (CT/i,1.X-4.X) scanners See Long Description for more Part #: 2154834 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.81 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: MA CONTROL BRD./CANCEL FRU: Yes Status: Orderable Comments: FLS - MA BOARD USED ON FAST RHAPSODE FMI#25238 UPGRADES THIS BD. TO A REV.1 ONE USE LIGHTSPEED ULTRA HQT WARNING!! MATERIAL MUST BE PURCHASED FROM MFGING DO NOT GO TO VENDOR AS IT MUST BE TESTED IN MFGING BEFORE SHIPPED TO FIELD VERY VERY HIGH DOA RATE. MS08992 GVR 7/31/00GELS
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.