Pular para o conteúdo principal
GE Healthcare · MRSigna2234211224151522415442241526

MiniGRAM undervoltage faults caused by MiniGRAM or MiniGRAM PS.

Symptom

MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image

Error message

2234211 MR Signal too large. Reduce Receive Gain 1. See Long Description Error: 2234211 Host: MR1SMR Process: NSP File: TPS:asc_check_RAP_overrange Line: 206 MR Signal too large; Reduce Receive Gain 1.

2241515 The GIP reported a Z Axis miniGRAM OverVoltage Fault. 2241515 HSC_MR01 NSP HPC_PORT_ID:GAP/GIP/GP 2092 The GIP reported a Z Axis miniGRAM OverVoltage Fault. Try swapping the miniGRAM Control Board, or the entire miniGRAM axis, or the miniGRAM Power Supply to isolate the problem.->

2241544 The gradient system on the Z axis recovered successfully. Mon Jun 25 23:45:29 2001 Error: 2241544 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 1342 The gradient system on the Z axis recovered successfully. There are no faults at the present time.

2241526 The miniGRAM Power Supply failed to go to Ready when commanded, or came out of Ready 2241526 HSC_MR01 NSP HPC_PORT_ID:GAP/GIP/GP 2481 The miniGRAM Power Supply failed to go to Ready when commanded, or came out of Ready. Possible causes: cable between GIP & miniGRAM Power Supply, the miniGRAM Power Supply, or the GIP.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2169727-4 MINI GRAM WITH FRU PACKAGING SIGNA ILLUSTRATED PARTS LIST - DIRECTION 2138247 REV 13 page 9–13 of GRADIENT CABINET section. 2169727–4 MINI GRAM WITH FRU PACKAGING SIGNA ILLUSTRATED PARTS LIST - DIRECTION 2138247 REV 13 page 9–13 of GRADIENT CABINET section. 2169727–4 MINI GRAM WITH FRU PACKAGING
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.