GE Healthcare · MRSigna2234211224151522415442241526
MiniGRAM undervoltage faults caused by MiniGRAM or MiniGRAM PS.
Symptom
MRIQ Collapsed Image Bright White Line Down or Across Middle of The Image
Error message
2234211 MR Signal too large. Reduce Receive Gain 1. See Long Description Error: 2234211 Host: MR1SMR Process: NSP File: TPS:asc_check_RAP_overrange Line: 206 MR Signal too large; Reduce Receive Gain 1. 2241515 The GIP reported a Z Axis miniGRAM OverVoltage Fault. 2241515 HSC_MR01 NSP HPC_PORT_ID:GAP/GIP/GP 2092 The GIP reported a Z Axis miniGRAM OverVoltage Fault. Try swapping the miniGRAM Control Board, or the entire miniGRAM axis, or the miniGRAM Power Supply to isolate the problem.-> 2241544 The gradient system on the Z axis recovered successfully. Mon Jun 25 23:45:29 2001 Error: 2241544 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP Line: 1342 The gradient system on the Z axis recovered successfully. There are no faults at the present time. 2241526 The miniGRAM Power Supply failed to go to Ready when commanded, or came out of Ready 2241526 HSC_MR01 NSP HPC_PORT_ID:GAP/GIP/GP 2481 The miniGRAM Power Supply failed to go to Ready when commanded, or came out of Ready. Possible causes: cable between GIP & miniGRAM Power Supply, the miniGRAM Power Supply, or the GIP.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2169727-4 MINI GRAM WITH FRU PACKAGING SIGNA ILLUSTRATED PARTS LIST - DIRECTION 2138247 REV 13 page 9–13 of GRADIENT CABINET section. 2169727–4 MINI GRAM WITH FRU PACKAGING SIGNA ILLUSTRATED PARTS LIST - DIRECTION 2138247 REV 13 page 9–13 of GRADIENT CABINET section. 2169727–4 MINI GRAM WITH FRU PACKAGING
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.