GE Healthcare · MRSigna2227966
AP failure due to memory module.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2227966 AP configuration error 0x2000. Possible problem with /usr/g/bin/prod.conf Error: 2227966 Host: lghmr Process: NSP File: TPS:ap init medcam Line: 398 AP configuration error 0x2000. Possible problem with /usr/g/bin/prod.conf
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2292662-3 128MB PMOD MEMORY 2292662-3 128MB PMOD MEMORY 2292662-3 128MB PMOD MEMORY
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.