GE Healthcare · MR2225520
TR driver Fault Body Path Short Due to CPD Board.
Symptom
Unable to Scan With T/R driver Fault, Body Path Short
Error message
2225520 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load. See Long Description Thu Sep 9 11:16:03 2004 Error: 2225520 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6080 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load.
Applies to
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Will not Scan with TR driver Body Path Short.
- T/R Driver Fault Due To a Shorted Body Hybrid Splitter.
- T/R Driver Fault Due To a Shorted Body Hybrid Splitter.
- Unable to Run Body RF Calibrations due to T/R Driver Faults due to SSM DRV FLT TR Switch Not in DIS Position.
- T/R Driver Faults Due To SSM.
- T/R Driver fault short caused by defective 50 ohm resistor in the Body Hybrid.