GE Healthcare · MROvationSigna2225520
T/R Driver fault short caused by defective 50 ohm resistor in the Body Hybrid.
Symptom
Unable to Scan With T/R driver Fault, Body Path Short
Error message
2225520 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load. See Long Description Thu Sep 9 11:16:03 2004 Error: 2225520 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6080 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load.
Applies to
- Models
- Ovation, Signa
- Product
- Ovation/MFO
- Software
- MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
- Component
- 2319498 .35T Teli Body Hybrid/TR Does not include body receive preamp nor 2279781 Bias Box (RF Switchbox)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.