Unable to Run Body RF Calibrations due to T/R Driver Faults due to SSM DRV FLT TR Switch Not in DIS Position.

Symptom

Unable to Run Body RF Calibrations due to T/R Driver Faults See Long Description The FE had just replaced the RF Amplifier and he was trying to run Body RF Output calibration but could not scan at all due the errors shown below. 2225519 T/R driver fault, Body Path open. Possibly a disconnected cable between the driver board and the body T/R network, or an open diode in the body T/R switch. 2225520 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load.

Error message

2225520 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load. See Long Description Thu Sep 9 11:16:03 2004 Error: 2225520 Host: oc001 Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6080 T/R driver fault, Body Path short. Possibly a shorted diode in the body T/R switch network, or a short circuit load.

2225519 T/R driver fault, Body Path open. Possibly a disconnected cable between the driver board and the body T/R network, or an open diode in the body T/R switch. 2225519 T/R driver fault, Body Path open. Possibly a disconnected cable between the driver board and the body T/R network, or an open diode in the body T/R switch.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.