The ConnectPro Worklist comes up empty when updated on the MR system.
Symptom
The ConnectPro Work List comes up empty when updated on the MR system see long description LX system running 9.1 M4 software. Just connected up to an INTERWORKS hisris Broker. The ConnectPro worklist comes up empty when queried from the MR. lx-mr 91# su - sdc {sdc@lx-mr}[1] getver Hostname : lx-mr Build number for MrpApps is 91.29I_M4_0311.b Build number for PostSdC is 91.29I_M4_0311.b Build number for cclass is 91.29I_M4_0311.b Build number for driverSupport is 91.29I_M4_0311.b Build number for insite is 91.29I_M4_0311.b Build number for install is 91.29I_M4_0311.b Build number for os_cd is 6_15.4.a Build number for patch_91_M4_02 is 91.29I_M4_0311.b.PB Build number for patch_91_M4_03 is 91.29I_M4_0311.b.PB Build number for patch_91_M4_05 is 91.29I_M4_0311.b.PJ Build number for patch_91_M4_07 is 91.29I_M4_0311.b.PG Build number for patch_91_M4_09 is 91.29I_M4_0311.b.PC Build number for patch_91_M4_11_0Connectivity is 91.29I_M4_0311.b.PK Build number for patch_91_M4_12 is 91.29I_M4_0311.b.PG Build number for patch_91_M4_14 is 91.29I_M4_0311.b.PJ Build number for patch_91_M4_15 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_16 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_17 is 91.29I_M4_0311.b.PL Build number for patch_91_M4_18 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_19 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_20 is 91.29I_M4_0311.b.PK Build number for pc is 91.29I_M4_0311.b MR Software release: 9.1.0311b
Error message
WLdestroyCodeType(): NULL worklist passed See Long Description Error seen in /usr/g/hisris/wlsErrors file: Fri 05-Mar-04 15:42:42 WLapiWorklist.c (331) Error code: 4 WLdestroyCodeType(): NULL worklist passed. 27 Zero length field/s found in the Worklist response Error seen in the /usr/g/hisris/wlsErrors file: Tue 09-Nov-04 14:04:35 WLS_mfm.c (1342) Error code: 27 Zero length field/s found in the Worklist response 34 Required Field, (Scheduled Station AE Title) (0040,0001) is zero length in the Worklist response Error seen in the /usr/g/hisris/wlsErrors file: Tue 09-Nov-04 14:04:35 WLServer.c (211) Error code: 34 Required Field, (Scheduled Station AE Title) (0040,0001) is zero length in the Worklist response WLserviceSignal(): Servicing the Kill Thu 17-Apr-:3 13:31:13 SrxHRIF.m (933) Error code: 0 WLserviceSignal(): Servicing the Kill 26 Warning Insufficient number of fields in the Worklist response Error seen in the /usr/g/hisris/wlsErrors file: Tue 09-Nov-04 14:04:35 WLS_mfm.c (1337) Error code: 26 :50Warning : Insufficient number of fields in the Worklist response
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.