The ConnectPro Worklist comes up empty when updated on the MR system.

Symptom

The ConnectPro Work List comes up empty when updated on the MR system see long description LX system running 9.1 M4 software. Just connected up to an INTERWORKS hisris Broker. The ConnectPro worklist comes up empty when queried from the MR. lx-mr 91# su - sdc {sdc@lx-mr}[1] getver Hostname : lx-mr Build number for MrpApps is 91.29I_M4_0311.b Build number for PostSdC is 91.29I_M4_0311.b Build number for cclass is 91.29I_M4_0311.b Build number for driverSupport is 91.29I_M4_0311.b Build number for insite is 91.29I_M4_0311.b Build number for install is 91.29I_M4_0311.b Build number for os_cd is 6_15.4.a Build number for patch_91_M4_02 is 91.29I_M4_0311.b.PB Build number for patch_91_M4_03 is 91.29I_M4_0311.b.PB Build number for patch_91_M4_05 is 91.29I_M4_0311.b.PJ Build number for patch_91_M4_07 is 91.29I_M4_0311.b.PG Build number for patch_91_M4_09 is 91.29I_M4_0311.b.PC Build number for patch_91_M4_11_0Connectivity is 91.29I_M4_0311.b.PK Build number for patch_91_M4_12 is 91.29I_M4_0311.b.PG Build number for patch_91_M4_14 is 91.29I_M4_0311.b.PJ Build number for patch_91_M4_15 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_16 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_17 is 91.29I_M4_0311.b.PL Build number for patch_91_M4_18 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_19 is 91.29I_M4_0311.b.PK Build number for patch_91_M4_20 is 91.29I_M4_0311.b.PK Build number for pc is 91.29I_M4_0311.b MR Software release: 9.1.0311b

Error message

WLdestroyCodeType(): NULL worklist passed See Long Description Error seen in /usr/g/hisris/wlsErrors file: Fri 05-Mar-04 15:42:42 WLapiWorklist.c (331) Error code: 4 WLdestroyCodeType(): NULL worklist passed.

27 Zero length field/s found in the Worklist response Error seen in the /usr/g/hisris/wlsErrors file: Tue 09-Nov-04 14:04:35 WLS_mfm.c (1342) Error code: 27 Zero length field/s found in the Worklist response

34 Required Field, (Scheduled Station AE Title) (0040,0001) is zero length in the Worklist response Error seen in the /usr/g/hisris/wlsErrors file: Tue 09-Nov-04 14:04:35 WLServer.c (211) Error code: 34 Required Field, (Scheduled Station AE Title) (0040,0001) is zero length in the Worklist response

WLserviceSignal(): Servicing the Kill Thu 17-Apr-:3 13:31:13 SrxHRIF.m (933) Error code: 0 WLserviceSignal(): Servicing the Kill

26 Warning Insufficient number of fields in the Worklist response Error seen in the /usr/g/hisris/wlsErrors file: Tue 09-Nov-04 14:04:35 WLS_mfm.c (1337) Error code: 26 :50Warning : Insufficient number of fields in the Worklist response

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.