GE Healthcare · CTLightSpeedVCT23001708239468023002304626000741416790642052189
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Fastcal fails the mylar window check. Fastcal fails the mylar window check
Error message
Error: 230017082 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.3 The row number can be VERY confusing. Please see Long Description for how to decode the ROW NUMBER Tue Mar 22 20:12:04 2005 Error: 230017082 Host: ct01 Process: accs File: aircal_processing.c Line: 917 Channel in non-bowtie airscan out of range - row 0, channel 10, value = 4.394680 Click Here to Decode the ROW NUMBER Error: 230023046 Calibration limit check failed. Beam may be obstructed See LD for complete error Tue Mar 22 20:12:04 2005 Error: 230023046 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3953 Calibration limit check failed. Beam may be obstructed Error: 260007414 Questionable DAS Channel Detected Mon Aug 15 07:34:52 2005 Error: 260007414 Host: DAS Process: DCB Firmware File: dcvtst.c Line: 202 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:34:52:365 Log Series: 12 Questionable DAS Channel Detected. DAS Output View Data did NOT Pass Selftest Converter Board Card: 44 Channel: 66. Channel Value Was: -32 Counts Expected: 1679064 to 2052189 Counts Average of All Channels Is: 1796721
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2336885 MDAS 4 or MDAS 8 Convertor Board, replaces 2322994-3. See Long Description for details. 2336885 MDAS 4 or MDAS 8 Convertor Card, replaces 2322994-3 (see below) Part #: 2336885 Part Type: GP CONSUMABLES Part Weight: 0.48 LBR Repairable: No Returnable: No Hazardous Material: N Description: SERVICE, MDAS CONVERTOR BD FRU: Yes Status: Orderable 4 Slice versus 8 Slice Configuration The HiSpeed QX/i system can operate 4 slice per rotation, data collections mode, whereas, the Detector and MDAS are capable of operating 4 slice or 8 slice per rotation. From a hardware perspective the main difference is in the MDAS. The 4 slice configuration uses a depopulated MDAS. This means that all EVEN slots are empty. Only ODD numbered slots contain converter boards. Data collection is still performed in the same manner using only the odd converter card slot assignments.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required