Pular para o conteúdo principal
GE Healthcare · CTLightSpeed200107023558911014433920050200510989

LightSpeed 4.X - Octane

Symptom

Images fail to display and recon shutdown. Images fail to display and recon shutdown

Error message

Error: 200107023 A Failure had been detected while calling UIR x_ToScanRequest function or srx_sm_generate_action function See Long Description for error variations Tue Oct 19 09:22:20 2004 Error: 200107023 Host: dgct Process: scanRx File: srx_sma_manage_scan.c Line: 688 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling UIRx_ToScanRequest function. Nov 22 08:55:38 2006 Error: 200107023 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 148 Function: Data Acquisition : OC Processing srx_sm_event_notify: A Failure had been detected while calling srx_sm_generate_action function. Tue Aug 14 16:28:25 2007 Error: 200107023 Host: cccgvct1 Process: scanRx File: srx_sma_manage_scan.c Line: 733 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling SIRx_areScanFilesAvailable function.

dksc0d1: [Alert] Media error : Write error (as seen in the Octane SYSLOG) See Long Description Example of a Octane disk drive WRITE Error, as seen in the /usr/adm/SYSLOG: cd /usr/adm more SYSLOG dksc0d1: <6>[Alert] Media error <6>: Write error (asc=0xc, asq=0x3)<6>, (data byte 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-81 Octane Computer 73 GB Disk Drive - replaces 2169940-78 (18GB) See long description for aditional information 2169940-81 Octane 73 GB Disk Drive, replaces 18 GB Disk 2169940-78 Part #: 2169940-81 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: CT OCTANE 2 73 GB DISK DRIVE FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.