Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260107404

LightSpeed 5.X (80/100kW)

Symptom

"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors

Error message

Error: 260107404 The Axial AC Drive reported fault code:F4- Undervoltage Fault. The Drive internal DC bus voltage dropped below 325V. Possible low voltage condition on 120VAC in gantry or power interruption. See long description for full error Thu Feb 22 18:00:09 2007 Error: 260107404 Host: tgp The Axial AC Drive reported fault code:F4- Undervoltage Fault. The Drive internal DC bus voltage dropped below 325V. Possible low voltage condition on 120VAC in gantry or power interruption.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2233402 Axial Drive Motor Controller See Long Description for details Part #: 2233402 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 8.64 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: AX DRIVE MTR CONTRLLER W/CORERS ASSY FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: MUST ORDER FROM HOLT ELECTRONICS. MS26561 DSM 1/7/01 USED ON HPOWER PROGRAM, PER RSPL. MS20544 ON 2/12/03. DO NOT DISCOUNT - PER SHAKUN*** 2ND OPINION NOTES *** Check fuses/connections associated with axial drive circuit; run axial functional diagnostic test and check for failures. If used on HP60
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues