HFO Z2 Resistive Shim Power Supply Software Installation Issues.
Symptom
HFO Z2 Resistive Shim Power Supply Software Installation Issues. See Long Description. Effectivity: All Signa 0.7T OpenSpeed LX Systems (HFO3 and earlier). Problem Description: After replacing a failed Resonance Research Z2 Shim Power Supply with a new Z2 Shim Power Supply (part number 2352972), the software fails to download from the Host to the Z2 Shim Power Supply when following the software load instructions in Direction 2407600. The procedures in sections 3-2 and 3-3 are missing steps critical to the success of the software download. SMILe 9/29/04
Applies to
- Models
- Signa
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.